Inter‐device agreement of sacral subepidermal oedema measurement in healthy adults during prolonged 60° head of bed elevation

Journal Publication ResearchOnline@JCU
Latimer, Sharon L.;Bone, Madeline;Walker, Rachel M.;Thalib, Lukman;Gillespie, Brigid M.
Abstract

Aim: To investigate the level of agreement between the SEM 200 and Provisio® subepidermal moisture sacral delta measurements, which may indicate increased pressure injury risk, in healthy adults during 120 min of prolonged 60° head of bed elevation. This position, which requires the elevation of the patient's upper body at a 60° angle above the horizontal plane for an extended period, is used by clinicians to prevent or manage a patient's medical or surgical conditions. Design: This prospective exploratory study recruited 20 healthy adults during October 2021 and collected sacral subepidermal moisture delta measurements using the SEM 200 and Provisio® devices. Methods: Delta measurements were taken at 20-min intervals over 120 min resulting in seven data collection timepoints. Descriptive statistics and a Bland Altman plot analysis were conducted. Results: A total of 280 sacral subepidermal moisture delta measurements were gathered or 140 per device. There were good levels of agreement between the two devices at baseline (T0) [mean 0.025; SD 0.137] and following 60- (T3) [mean 0.025; SD 0.111], 80- (T4) [mean −0.01; SD 0.177] and 100 min (T5) [mean 0.01; SD 0.129] of prolonged 60° head of bed elevation. Head of bed elevations can increase a patient's risk of sacral pressure injuries. In some countries, nurses have access to the SEM 200 and/or the Provisio® device, so our findings may increase nurses' confidence in the interchangeability of the device measurements, although further research is needed to confirm this. The SEM 200 and Provisio® subepidermal moisture scanners show promise in gathering similar objective pressure injury risk data which could prompt clinicians to implement prevention strategies. Impact: Current pressure injury risk assessment is largely subjective in nature. This quantitative study on healthy human sacral tissue found a good level of agreement in the SEM 200 and Provisio® subepidermal moisture scanners, which may increase nurses' confidence in the interchangeability of the devices in clinical practice.

Journal

Nursing Open

Publication Name

N/A

Volume

11

ISBN/ISSN

2054-1058

Edition

N/A

Issue

2

Pages Count

8

Location

N/A

Publisher

Wiley

Publisher Url

N/A

Publisher Location

N/A

Publish Date

N/A

Url

N/A

Date

N/A

EISSN

N/A

DOI

10.1002/nop2.2103