Real-Time Temperature Sensing Using a Ratiometric Dual Fluorescent Protein Biosensor

Journal Publication ResearchOnline@JCU
Sorenson, Alanna E.;Schaeffer, Patrick M.
Abstract

Accurate temperature control within biological and chemical reaction samples and instrument calibration are essential to the diagnostic, pharmaceutical and chemical industries. This is particularly challenging for microlitre-scale reactions typically used in real-time PCR applications and differential scanning fluorometry. Here, we describe the development of a simple, inexpensive ratiometric dual fluorescent protein temperature biosensor (DFPTB). A combination of cycle three green fluorescent protein and a monomeric red fluorescent protein enabled the quantification of relative temperature changes and the identification of temperature discrepancies across a wide temperature range of 4–70 °C. The maximal sensitivity of 6.7% °C−1 and precision of 0.1 °C were achieved in a biologically relevant temperature range of 25–42 °C in standard phosphate-buffered saline conditions at a pH of 7.2. Good temperature sensitivity was achieved in a variety of biological buffers and pH ranging from 4.8 to 9.1. The DFPTB can be used in either purified or mixed bacteria-encapsulated formats, paving the way for in vitro and in vivo applications for topologically precise temperature measurements.

Journal

Biosensors

Publication Name

N/A

Volume

13

ISBN/ISSN

2079-6374

Edition

N/A

Issue

3

Pages Count

11

Location

N/A

Publisher

MDPI

Publisher Url

N/A

Publisher Location

N/A

Publish Date

N/A

Url

N/A

Date

N/A

EISSN

N/A

DOI

10.3390/bios13030338